Extracorporeal life support

From IDWiki
Revision as of 00:59, 15 June 2021 by Aidan (talk | contribs) (Created page with "* Cannulation ** VV provides respiratory support but no circulatory support ** VA provides both respiratory and circulatory support * Extracorporeal circuit ** Blood flow: sta...")
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
  • Cannulation
    • VV provides respiratory support but no circulatory support
    • VA provides both respiratory and circulatory support
  • Extracorporeal circuit
    • Blood flow: start with 50-80 mL/kg/min, titrating to lowest flow to maintain SaO2 >80-85% at rest vent settings
    • Oxygenation: usually targets SaO2 80-85%
    • Sweep flow: typically 1:1 gas:blood; increased flow will increase CO2 clearance without affecting oxygenation
      • e.g. for purely hypercapneic respiratory failure, blood flow can be very low (0.5 L/min/m2) while sweep flow is high (10:1)
      • Titrate to maintain PaCO2 at 40 mmHg
    • Sweep gas: typically 95 to 100% FiO2