== Background == |
|||
== Disinfection & Sterilization == |
|||
=== Definitions === |
|||
* Equipment must be ''cleaned'' first, before attempting to disinfect or sterilize it |
|||
* '''Cleaning''' is the process of removing soil and organic material, usually with water and detergents or enzymes |
|||
=== Spaulding Classification of Medical Equipment === |
|||
* '''Sterilization''' is a process that eliminates all forms of microbial life |
|||
* '''Disinfection''' is a process that eliminates many or all pathogenic microbes on inanimate objects, with the exception of spores |
|||
* '''Chemical sterilants''' are disinfectants that will kill spores with prolonged exposure (2 to 12 hours) |
|||
* '''High-level disinfectants''' will kill all microorganisms except large numbers of bacterial spores within a shorter amount of time (e.g. 20 minutes) |
|||
* '''Low-level disinfectants''' will kill most vegetative bacteria, some fungi, and some viruses within a short amount of time (e.g. 10 minutes) |
|||
* '''Intermediate-level disinfectants''' may be cidal for mycobacteria, vegetative bacteria, most viruses, and most fungi, but are not sporicidal |
|||
=== Sterilization === |
|||
* The process of sterilization destroys all forms of microbial life |
|||
*May be physical or chemical methods |
|||
**Steam under pressure |
|||
**Dry heat |
|||
**EtO gas |
|||
**Hydrogen peroxide gas plasma |
|||
**Liquid chemicals |
|||
*Equipment must be ''cleaned'' first, before attempting to disinfect or sterilize it |
|||
== Spaulding Classification of Medical Equipment == |
|||
{| class="wikitable" |
{| class="wikitable" |
||
Line 24: | Line 42: | ||
| Intact skin |
| Intact skin |
||
| Cleaning followed by low level disinfection |
| Cleaning followed by low level disinfection |
||
| |
| Tourniquets, BP cuffs, linen, and furniture |
||
|} |
|} |
||
=== Rutala Modification === |
|||
* Endoscopes, which secondarily enter sterile tissues, should be considered critical |
|||
== Methods of Sterilization and Disinfection == |
|||
{| class="wikitable" |
|||
! rowspan="2" |Method |
|||
! rowspan="2" |Corrodes Metal |
|||
! colspan="4" |Smooth, Hard Surfaces |
|||
! colspan="2" |Rubber Tubing and Catheters |
|||
! colspan="2" |Polyethylene Tubing and Catheters |
|||
! colspan="2" |Lensed Instruments |
|||
!Thermometers |
|||
! colspan="2" |Hinged Instruments |
|||
|- |
|||
!S |
|||
!HLD |
|||
!ILD |
|||
!LLD |
|||
!S |
|||
!HLD |
|||
!S |
|||
!HLD |
|||
!S |
|||
!HLD |
|||
!LLD |
|||
!S |
|||
!HLD |
|||
|- |
|||
|Heat sterilization, including steam or hot air for 3 to 30 min |
|||
| |
|||
|X |
|||
| |
|||
| |
|||
| |
|||
|X |
|||
| |
|||
|X |
|||
| |
|||
|X |
|||
| |
|||
| |
|||
|X |
|||
| |
|||
|- |
|||
|Ethylene oxide gas for 1 to 6 hours |
|||
| |
|||
|X |
|||
| |
|||
| |
|||
| |
|||
|X |
|||
| |
|||
|X |
|||
| |
|||
|X |
|||
| |
|||
| |
|||
|X |
|||
| |
|||
|- |
|||
|Hydrogen peroxide gas plasma for 45 to 72 minutes |
|||
| |
|||
|X |
|||
| |
|||
| |
|||
| |
|||
|X |
|||
| |
|||
|X |
|||
| |
|||
|X |
|||
| |
|||
| |
|||
|X |
|||
| |
|||
|- |
|||
|Glutaraldehyde-based formulations (>2% gluteraldehyde) or gluteraldehyde 1.12% and phenol/phenate 1.93% |
|||
| |
|||
|10 h |
|||
|X |
|||
| |
|||
| |
|||
|10 h |
|||
|X |
|||
|10 h |
|||
|X |
|||
|10 h |
|||
|X |
|||
| |
|||
|10 h |
|||
|X |
|||
|- |
|||
|Hydrogen peroxide 7.5% |
|||
|X |
|||
|6 h |
|||
|X |
|||
| |
|||
| |
|||
|6 h |
|||
|X |
|||
|6 h |
|||
|X |
|||
|6 h |
|||
|X |
|||
| |
|||
|6 h |
|||
|X |
|||
|- |
|||
|Peracetic acid >0.2% for 12 min at 50 to 56ºC |
|||
| |
|||
|X |
|||
| |
|||
| |
|||
| |
|||
|X |
|||
| |
|||
|X |
|||
| |
|||
|X |
|||
| |
|||
| |
|||
|X |
|||
| |
|||
|- |
|||
|Hydrogen peroxide 7.35% and peracetic acid 0.23%, or hydrogen peroxide 1% and peracetic acid 0.08% |
|||
|X |
|||
|3 to 8 h |
|||
|X |
|||
| |
|||
| |
|||
|3 to 8 h |
|||
|X |
|||
|X |
|||
|X |
|||
|3 to 8 h |
|||
|X |
|||
| |
|||
|X |
|||
|X |
|||
|- |
|||
|Ortho-phthalaldehyde (OPA) 0.55% |
|||
| |
|||
| |
|||
|X |
|||
| |
|||
| |
|||
| |
|||
|X |
|||
| |
|||
|X |
|||
| |
|||
|X |
|||
| |
|||
| |
|||
|X |
|||
|- |
|||
|Wet pasteurization (30 min at 70ºC) with detergent cleaning |
|||
| |
|||
| |
|||
|X |
|||
| |
|||
| |
|||
| |
|||
|X |
|||
| |
|||
|X |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|X |
|||
|- |
|||
|Hypochlorite generated by electrolyzing saline |
|||
|X |
|||
| |
|||
|X |
|||
| |
|||
| |
|||
| |
|||
|X |
|||
| |
|||
|X |
|||
| |
|||
|X |
|||
| |
|||
| |
|||
|X |
|||
|- |
|||
|Ethyl or isopropyl alcohol 70 to 90% |
|||
| |
|||
| |
|||
| |
|||
|X |
|||
|X |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|X |
|||
| |
|||
| |
|||
|- |
|||
|Sodium hypochlorite 5.25 to 6.15% (diluted household bleach) |
|||
| |
|||
| |
|||
| |
|||
|X |
|||
|X |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|- |
|||
|Phenolic germicidal detergent solution |
|||
| |
|||
| |
|||
| |
|||
|X |
|||
|X |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|- |
|||
|Iodophor germicidal detergent solution |
|||
| |
|||
| |
|||
| |
|||
|X |
|||
|X |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|- |
|||
|Quaternary ammonium germicidal detergent solution |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|X |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|} |
|||
* Exposure times: |
|||
* Endoscopes, which secondarily enter steril tissues, should be considered critical |
|||
** Sterilization: varies by method |
|||
** High-level disinfection: 12 to 30 minutes at ≥20ºC |
|||
** Intermediate disinfection: ≥1 min |
|||
** Low-level disinfection: ≥1 min |
|||
== Medical Device Reprocessing Department (MDRD) == |
|||
* Needs separate areas for receiving soiled instruments, decontaminating, cleaning, sterilizing, and storing |
* Needs separate areas for receiving soiled instruments, decontaminating, cleaning, sterilizing, and storing |
||
Line 37: | Line 327: | ||
* Negative pressure in decontamination areas and positive pressure in clean areas |
* Negative pressure in decontamination areas and positive pressure in clean areas |
||
=== Pre-Cleaning === |
|||
* Dismantle and sort devices, then soak them |
* Dismantle and sort devices, then soak them |
||
=== Cleaning === |
|||
* Remove all visible soil from every accessible surface |
* Remove all visible soil from every accessible surface |
||
Line 47: | Line 337: | ||
* Rinse to remove detergents, then dry |
* Rinse to remove detergents, then dry |
||
=== Sterilization === |
|||
* High temperature sterilization whenever possible |
* High temperature sterilization whenever possible |
||
Line 56: | Line 346: | ||
** Hydrogen peroxide most common liquid |
** Hydrogen peroxide most common liquid |
||
==== Monitoring ==== |
|||
* Spore tests: compare a strip that went through the autoclav to a control that was held outside |
* Spore tests: compare a strip that went through the autoclav to a control that was held outside |
||
* ATP system test strips |
* ATP system test strips |
||
=== Storage === |
|||
* Equipment must be stored until results are available from the spore test for that batch |
* Equipment must be stored until results are available from the spore test for that batch |
||
* Controlled temperature and humidity |
* Controlled temperature and humidity |
||
== Further Reading == |
|||
* [https://apic.org/ APIC] |
* [https://apic.org/ APIC] |
Latest revision as of 18:17, 25 October 2021
Background
Definitions
- Cleaning is the process of removing soil and organic material, usually with water and detergents or enzymes
- Sterilization is a process that eliminates all forms of microbial life
- Disinfection is a process that eliminates many or all pathogenic microbes on inanimate objects, with the exception of spores
- Chemical sterilants are disinfectants that will kill spores with prolonged exposure (2 to 12 hours)
- High-level disinfectants will kill all microorganisms except large numbers of bacterial spores within a shorter amount of time (e.g. 20 minutes)
- Low-level disinfectants will kill most vegetative bacteria, some fungi, and some viruses within a short amount of time (e.g. 10 minutes)
- Intermediate-level disinfectants may be cidal for mycobacteria, vegetative bacteria, most viruses, and most fungi, but are not sporicidal
Sterilization
- The process of sterilization destroys all forms of microbial life
- May be physical or chemical methods
- Steam under pressure
- Dry heat
- EtO gas
- Hydrogen peroxide gas plasma
- Liquid chemicals
- Equipment must be cleaned first, before attempting to disinfect or sterilize it
Spaulding Classification of Medical Equipment
Classification | Contact with | Processing | Examples |
---|---|---|---|
Critical | Sterile tissue or vascular system | Cleaning followed by sterilization | Surgical instruments, biopsy instruments, and foot care equipment |
Semi-critical | Non-intact skin or mucous memranes, but not penetrating the skin | Cleaning followed by a minimum high level disinfection | Respiratory equipment, flexible endoscopes |
Non-critical | Intact skin | Cleaning followed by low level disinfection | Tourniquets, BP cuffs, linen, and furniture |
Rutala Modification
- Endoscopes, which secondarily enter sterile tissues, should be considered critical
Methods of Sterilization and Disinfection
Method | Corrodes Metal | Smooth, Hard Surfaces | Rubber Tubing and Catheters | Polyethylene Tubing and Catheters | Lensed Instruments | Thermometers | Hinged Instruments | |||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
S | HLD | ILD | LLD | S | HLD | S | HLD | S | HLD | LLD | S | HLD | ||
Heat sterilization, including steam or hot air for 3 to 30 min | X | X | X | X | X | |||||||||
Ethylene oxide gas for 1 to 6 hours | X | X | X | X | X | |||||||||
Hydrogen peroxide gas plasma for 45 to 72 minutes | X | X | X | X | X | |||||||||
Glutaraldehyde-based formulations (>2% gluteraldehyde) or gluteraldehyde 1.12% and phenol/phenate 1.93% | 10 h | X | 10 h | X | 10 h | X | 10 h | X | 10 h | X | ||||
Hydrogen peroxide 7.5% | X | 6 h | X | 6 h | X | 6 h | X | 6 h | X | 6 h | X | |||
Peracetic acid >0.2% for 12 min at 50 to 56ºC | X | X | X | X | X | |||||||||
Hydrogen peroxide 7.35% and peracetic acid 0.23%, or hydrogen peroxide 1% and peracetic acid 0.08% | X | 3 to 8 h | X | 3 to 8 h | X | X | X | 3 to 8 h | X | X | X | |||
Ortho-phthalaldehyde (OPA) 0.55% | X | X | X | X | X | |||||||||
Wet pasteurization (30 min at 70ºC) with detergent cleaning | X | X | X | X | ||||||||||
Hypochlorite generated by electrolyzing saline | X | X | X | X | X | X | ||||||||
Ethyl or isopropyl alcohol 70 to 90% | X | X | X | |||||||||||
Sodium hypochlorite 5.25 to 6.15% (diluted household bleach) | X | X | ||||||||||||
Phenolic germicidal detergent solution | X | X | ||||||||||||
Iodophor germicidal detergent solution | X | X | ||||||||||||
Quaternary ammonium germicidal detergent solution | X |
- Exposure times:
- Sterilization: varies by method
- High-level disinfection: 12 to 30 minutes at ≥20ºC
- Intermediate disinfection: ≥1 min
- Low-level disinfection: ≥1 min
Medical Device Reprocessing Department (MDRD)
- Needs separate areas for receiving soiled instruments, decontaminating, cleaning, sterilizing, and storing
- Treated essentially like an OR, with controlled temperature and humidity, with dedicated ventilation system
- Negative pressure in decontamination areas and positive pressure in clean areas
Pre-Cleaning
- Dismantle and sort devices, then soak them
Cleaning
- Remove all visible soil from every accessible surface
- Can use ultrasonic machines and washing machines
- Rinse to remove detergents, then dry
Sterilization
- High temperature sterilization whenever possible
- Low-temperature sterilization can be done for heat sensitive critical and semi-critical items
- Uses gas, including hydrogen peroxide gas plasma or vapour, or ozone (but that needs 4 hours)
- Needed for things with plastic, for example
- Liquid emersion can be done for heat sensitive crticical and semi-critical items
- Hydrogen peroxide most common liquid
Monitoring
- Spore tests: compare a strip that went through the autoclav to a control that was held outside
- ATP system test strips
Storage
- Equipment must be stored until results are available from the spore test for that batch
- Controlled temperature and humidity